S IMPLIFIBER ENGLISH - 41
- Upload the stored test results to your PC.
- Delete saved results from the units memory.
Err 4 - No results are stored in
memory.
Error code 4 indicates that there are no results
stored in the units memory.
- SIMPLIFIBERs memory is empty.
Err 5 - Wavelength cannot be
changed.
Error code 5 indicates that the wavelength
cannot be changed while the SIMPLIFIBER Source
unit is in AUTO mode.
- Change mode to CW prior to changing the
wavelength
Err 6 - Calibration is expired.
Calibration values are corrupted
- Send unit in for calibration
Err 7 - Communication error be-
tween SIMPLIFIBER Meter and PC.
Communication between PC and SIMPLIFIBER
cannot be established.
- Verify connection between SIMPLIFIBER and the
PC and retry.