MESSAGE CAUSE WHAT TO DO
Forced to stop the blood
pressure measurement after
repump for 2 times.
Turn off the device and
start over. Follow the
instructions and try again.
Data corrupted.
Follow the instructions
and try again. If problem
persists, contact the local
customer service for as-
sistance.
Strip has been used.
Repeat the test with a new
strip.
Problem in operation.
Review the instructions
and repeat the test with a
new strip.
Environmental temperature
is outside the system’s
operational range.
System operational range
is 10°C to 4C (50°F to
104°F). Repeat the test
after the monitor and test
strip have reached the
above temperature.