45
Dimension & Weight: 90.7mm(L) x 46.3 mm(W) x 21.5 mm(H), 57.9g
Power source: two 1.5V AAA alkaline batteries
Battery life: Over 1,000 determinants
Display: Large LCD
Memory: 450 measurement results with respective dates and times
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time countdown
Sleeping mode: Power consumption less than 50uA
Auto shutdown after 3 minutes of idleness
Temperature warning
Operating condition: 50°F to104°F (10°C to 40°C), below 85% R.H.
(noncondensing)
Meter storage/transportation condition:-4°F to 140°F (-20°C to 60°C),
below 95% R.H.
Shelf life: 3 years
Measurement units: mg/dL
Measurement range: 20 to 600 mg/dL(1.1 to 33.3 mmol/L)
The device has been tested to meet the electrical and safety requirements of:
IEC/EN 61010-1, IEC/EN 61010-2-101 & EN 61326.
Performance characteristics
Accuracy
Within ±15 mg/dL (0.83 mmol/L) at glucose concentration < 75 mg/dL (4.2
mmol/L) and within ±20% at glucose concentration ≥ 75 mg/dL (4.2 mmol/L).
Precision
CVs (%) of intermediate precision and repeatability are less than 5%.
The product has been tested to meet the requirements of ISO 15197.
SPECIFICATIONS