22
Fully insert the test strip, with the contact bars end rst facing up
into test slot. The meter will turn on automatically. “CH” and “
” will appear rst, followed by a ashing “
”symbol.
After the symbol appears on display,
press the M button so that “CTL” appears.
With the “CTL” sign on the display, the
meter will not store your test result in
memory. If you decide not to perform a
control solution test, press the M button
again and the “CTL” sign will disappear.
STEP 1 Insert Test Strip
STEP 2 Press the M Button
Performing a Control Solution Test
PLEASE NOTE
1. Contact bars must be inserted all the way into the meter or you may
receive an inaccurate test result.
2. Every time you perform a control solution test, you must enter into the
“CTL” test mode so that the test result will not be stored in the meter
memory. Failure to do so will cause confusion between the blood glu-
cose test result and the control solution test result in system memory.