The NT8 has humidifiers that actively maintain moisture levels up to 90% to prevent drop
evaporation. Please ensure that the NT8 chambers have enough water for proper function.
To configure your humidifiers:
1. Click the Options button and go to the Advanced Options tab.
2. Configure the Humidifier Control settings.
● Enabled: If set to True, enables the NT8 to monitor and control the humidifiers
as necessary.
● Humidity Level (%): Set the humidity level to turn off the humidifier. When the
humidity level reaches the limit, the system automatically turns off the humidifier.
3. Click Save.
Occasionally you might want to empty and clean the system liquid and humidifier chamber
that serve your NT8. To do so, you'll need to turn the pump control settings off while you're
cleaning the chambers and then turn them back on.
Prerequisites
● Humidifier is on manual mode (click on the toolbar to switch to manual mode).
Emptying and Filling System Liquid and Humidifier Chambers
Important: The following tutorial is for advanced users and administrators only.
Important: The following tutorial is for advanced users and administrators only.
M