NT8 User's Guide | Chapter 15: System Maintenance
The pump will run until the fluid drains to the minimum level and begin to fill the
humidifier reservoir with the distilled water until it reaches the maximum fluid
level.
Repeat this process two additional times for a total of three drain/fill cycles.
9. Rinse the NT8’s liquid lines with the distilled water:
Click the Run Task List button to run the task list.
When the NT8 finishes running the task list, wait for 5 seconds to let the pump
rest in order to prevent overheating.
Repeat this process two additional times for a total of three rinse cycles.
Your NT8 is now sterilized and ready for dispensing.
Lubricating the plate copy plungers is an important maintenance step that should be done
every 6 months or as needed. If you are noticing aspiration or dispense issues when trying to
perform a plate copy (also known as creating daughter plates from a deep well block), you
should perform the actions in this tutorial as a troubleshooting step. The estimated time
required to lubricate the plate copy plungers is 30 minutes.
Before you begin, please make sure there are no Low Volume tips attached to the NT8's head.
You will need:
The hex keys (shipped in your accessory kit).
A small paint brush (shipped in your accessory kit).
Grease (shipped in your accessory kit).
A deep well block filled about half way with water colored with food dye.
A daughter plate.
A High Volume tip caddy with 8 HV tips.
The PCTesting.msq script file (starting with NT8 v2.5, this file is located at
C:\ProgramData\Formulatrix\NT8\{serial_number}\Data\User\Scri
pts. If not, email support@formulatrix.com and request the file).
Access to the Programmatic Tab.
Lubricating the Plate Copy Plungers