59
NT8 User's Guide | Chapter 13: LCP Experiments
Next to Z-Offset, enter the distance from the plate you want the pipette tips to
dispense from.
OR
Click the FF button to dispense directly to the plate surface.
10. Click End Repeat in the Task List to select it.
11. Double-click the Waste Station image to add the Dispose Tip command to the Task
List.
Run the LCP Dispense
Click the Run Task List button to execute the Task List. The NT8 picks up the LV
tips, washes them, aspirates from the microplate containing the well solution, then
dispenses the LCP sample across a column and then dispenses a column of well
solution to the LCP sample drops. This process repeats for each column you specified
for the LCP plate. After the NT8 has completed the tasks for each column, it disposes
the tips in the waste station.
Attach the LCP Coverslip
Gently place the coverslip on the dispensed LCP plate. Use a roller or spatula to apply
force to the coverslip to affix the coverslip to the base plate.