5 Maintenance
It is essential that your FOSS instrument receives a certain amount of regular
maintenance to ensure optimum performance. This in turn ensures quality and
consistency of results at the highest level.
To keep the Cyclotec Sample Mill clean, it is recommended to clean the grinding
chamber and dust pad every time finish grinding samples. Please use brush to clean
the impeller and mild detergent to clean the mill.
5.1 Cleaning of Dust Pad
The dust pad has a high efficiency and needs to be cleaned when the grinding
capacity becomes lower. This is done according to
Fig. 5:1.
Fig. 5:1
1. Remove the dust pad from the top of the sample outlet.
2. Shake the dust pad to remove the dust.
3. If the dust pad is quite dirty, wash it with cleaning water. Then dry it at 103°C in
an oven.
4. Replace in its original place.
60070581009a