6 Troubleshooting
6.1 Diagnostics and Troubleshooting
Fault Probable Cause Action
Cannot start instrument
Mains cord disconnected. Connect mains cord.
Thermo fuse not activate. Reset thermo fuse.
Grinding chamber cover
or/and sample inlet
chamber are not closed.
Close grinding and
sample inlet chambers.
Motor failure.
Contact the FOSS Service
Department.
Instrument stops during
the operation
High current. Reset thermo fuse.
Impeller stuck or too high
sample inlet speed.
Switch off and check if
impeller is stuck; reduce
sample inlet speed.
Motor or brake failure.
Contact the FOSS Service
Department.
Belt jumps or slips
Spring of tensioning
pulley is too loose or
broken.
Contact the FOSS Service
Department.