Original manual – Iss. 0.1 User manual – BM 3E Freezer 10/ 44
4.5 Performance and ambient temperature
The BM freezer is factory set to optimise its electricity consumption.
On the -86°C model, the setpoint can be adjusted from -55°C to -87°C. The default setting is -80°C.
On the -45°C model, the setpoint can be adjusted from -20°C to -45°C. The default setting is -45°C.
The ambient temperature (between +18°C and +35°C, ideally between 23°C and 25°C) also has a significant impact on
the electricity consumption of the equipment.