EVO EXPERT Manual 22/105
On this screen, the progress bar that appears below the word ¨CALIBRATING¨ displays the status
of this process accompanied by a textual representation of the progress in a percent format.
The exit key ‘MEM/CLEAR’ allow us to exit to the main menu but never while calibrating (never
while the screen looks like the example just above).
NOTE: Exiting mid-calibration denies the equipment a proper calibration and
therefore it cannot guarantee accurate results.
When the calibration process is over, information on the values of the angles and curvatures of the
calibration are displayed, as it is shown in the following screen:
If the curvature is lower to 2%, hit ‘ENTER’ to confirm the calibration and you will be taken to the
following screen:
Once on this screen, select the ‘SET CALIBRATION’ option using the and keys and then press
‘ENTER’. The calibration performed will be stored permanently in the viscometer’s memory. Then,
the viscometer will show the main menu screen. The Factory Calibration is now completed and it
can be restored as the default calibration at any moment.
The procedure for the User Calibration is similar to the above. From the ‘CALIBRATION’ menu select
the ‘USER CALIBRATION’ option. The following screen will appear: