EVO EXPERT Manual 23/105
Select the ‘VISCOSITY’ option and you will access to the following screens, depending on the model
of your viscometer:
Model L
Models R and H
Upon entering this screen, the spindle field is highlighted in light blue background. Using the ‘ and
’ keys you can change the Spindle.
The list of possible spindles to use depends on the model of your viscometer (L, R or H). Thus, in
tables 8 through 17 (page 62 and on) you can see the different spindles available for each model.
Note: It is recommended to use the following combinations of Spindle and standard
oil depending on the viscometer model:
Model L: TL5 Spindle (LCP adapter) and RT-50 (50 cP) standard
Model R: TR8 Spindle (LCP adapter) and RT-500 (500 cP) standard
Model H: TR8 Spindle (LCP adapter) and RT-500 (500 cP) standard
Once you’ve selected your spindle press ‘ENTER’ and the ‘VISCOSITY’ field will be highlighted. Press
‘ENTER’ again and the following screen will appear: