Table 13. EVO R Special spindle selection
Maximum guideline values in cP (mPa·s)
RPM / SP
TR8
TR9
TR10
TR11
0.01
5M
25M
50M
100M
0.3
166.6K
833.3K
1.6M
3.3M
0.5
100K
500K
1M
2M
0.6
83.3K
416.6K
833.3K
1.6M
1
50K
250K
500K
1M
1.5
33.3K
166.6K
333.3K
666.6K
2
25K
125K
250K
500K
2.5
20K
100K
200K
400K
3
16.6K
83.3K
166.6K
333.3K
4
12,5K
62,5K
125K
250K
5
10K
50K
100K
200K
6
8.3K
41.6K
83.3K
166.6K
10
5K
25K
50K
100K
12
4.16K
20.83K
41.6K
83.3K
20
2.5K
12,5K
25K
50K
30
1.6K
8.3K
16.6K
33.3K
50
1K
5K
10K
20K
60
833.3
4.16K
8.3K
16.6K
100
500
2.5K
5K
10K
200
250
1.25K
2.5K
5K
250
200
1K
2K
4K
ATTENTION:
K Indicates miles. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of
the selected scale.