Table 14. LCP Adaptor for EVO R
Maximum guideline values in cP (mPa·s)
RPM
LCP
0.01
640000.00
0.3
21333.00
0.5
12800.00
0.6
10666.00
1
6400.00
1.5
4266.00
2
3200.00
2.5
2560.00
3
2133.00
4
1600.00
5
1280.00
6
1066.00
10
640.00
12
533.00
20
320.00
30
213.00
50
128.00
60
106.00
100
64.00
200
32.00
250
27.00
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of the
selected scale
Volume of the sample = 16 ml.
Shear Rate = 1.2236·rpm