Table 17. HELDAL special spindle selection for EVO L
Maximum guideline values in cP (mPa·s)
RPM/SP
PA
PB
PC
PD
PE
PF
0.3
62.4K
124.8K
312K
624K
1.56M
3.12M
0.5
37.44K
74.88K
187.2K
374.4K
936K
1.872M
0.6
31.2K
62.4K
156K
312K
780K
1M
1
18.72K
37.44K
93.6K
187.2K
468K
936K
1.5
12.48K
24.96K
62.4K
124.8K
312K
624K
2
9.36K
18.72K
46.8K
93.6K
234K
468K
2.5
7.488K
14.976K
37.44K
74.88K
187.2K
374.4K
3
6.24K
12.48K
31.2K
62.4K
156K
312K
4
4.68K
9.36K
23.4K
46.8K
117K
234K
5
3.744K
7.488K
18.72K
37.44K
93.6K
187.2K
6
3.120K
6.24K
15.6K
31.2K
78K
156K
10
1.872K
3.744K
9.36K
18.72K
46.8K
93.6K
12
1.560K
3.12K
7.8K
15.6K
39K
78K
ATTENTION:
K Indicates miles. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of the
selected scale.