Table 18. HELDAL special spindle selection for EVO R
Maximum guideline values in cP (mPa·s)
RPM/SP
PA
PB
PC
PD
PE
PF
0.01
20M
40M
100M
200M
500M
1000M
0.3
666.6K
1.3M
3.3M
6.6M
16.6M
33.3M
0.5
400K
800K
2M
4M
10M
20M
0.6
333.3K
666.6K
1.6M
3.3M
8.3M
16.6M
1
200K
400K
1M
2M
5M
10M
1.5
133.3K
266.6K
666.6K
1.3M
3.3M
6.6M
2
100K
200K
500K
1M
2.5M
5M
2.5
80K
160K
400K
800K
2M
4M
3
66.6K
133.3K
333.3K
666.6K
1.6M
3.3M
4
50K
100K
250K
500K
1.25M
2.5M
5
40K
80K
200K
400K
1M
2M
6
33.3K
66.6K
166.6K
333.3K
833.3K
1.6M
10
20K
40K
100K
200K
500K
1M
12
16.6K
33.3K
83.3K
166.6K
416.6K
833.2K
ATTENTION:
K Indicates miles. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of the
selected scale.