Chapter 3 -- Installation--Firmware Update
3 - 13
Figure 3-9
Firmware Update Tab
Depending upon which code you wish to update, click on Update Instrument Firmware, Update
Communications Firmware, or Update PLD Firmware. You will then be prompted for a file. Navigate to the
file’s location (on the Gamry software CD or on your computer’s hard drive) and then press Open. The
update procedure will begin. A status bar shows the progress of the update. The USB indicator on the
Interface 1000 should also turn red during the procedure. Once the update is successful, press OK and your
Interface 1000 will be ready for use.
Once again, appropriate update files can be obtained from the Gamry Instruments website at
www.gamry.com. Should you encounter a problem updating the firmware in your Interface 1000, please
contact Gamry Instruments for assistance.
Interrupting a firmware update can cause a catastrophic failure of your system. Do not turn off the Interface
1000, do not unplug the USB cable, and do not stop the host computer operation when the USB LED is a
steady red color.
Caution
Do not interrupt a firmware update that is in progress. An incomplete update can
render an Interface 1000 inoperable until it is returned to Gamry for reprogramming.