Appendix A -- Interface 1000 Specifications--Floating Operation
9 - 3
Potentiostatic Mode
Applied Voltage Range Min
± 12
volts
Accuracy
DC zero offset
Gain
Max
1
0.2
mV
% setting
Note 18
DC Bias Typ
± 8
volts
Scan DAC ranges Typ
± 6.4, ± 1.6, ± 0.4
volts
Drift Max < 20
µV/°C
Note 19
Noise and Ripple
1 Hz to 1 kHz
1 Hz to 200 kHz
Typ
10
20
µV rms
Note 20
Galvanostatic Mode
Maximum Full Scale Current
± 1000
mA
Minimum Full Scale Current 10 nA
Accuracy Dominated by current to voltage
converter accuracy shown above
Sig Gen Voltage for Full Scale Current
± 3.0 volts
Environmental
Operating Temperature Range 0 to 45
°C
Relative Humidity Max 90 (non-condensing) %
Storage and Shipping Temperature -25 to 75
°C
Maximum Shipping acceleration 30 G