Environmental conditions
DataParameter
18°C to 28°CTemperatureOperating conditions
20% to 70%, RH (no dew condensa-
tion)
Humidity
-20°C to 70°CTemperatureTransportation and stor-
age temperature
5% to 95% (no dew condensation)Humidity
It is recommended to lower the lighting level to about
20 lux when moving a sample from the cassette into
Amersham Typhoon after exposure.
Lighting
Indoor use only, out of direct sunlight or brightly lit sur-
roundings.
Where to use
2000 m above sea levelMax. altitude, operation
Transient overvoltage category IIOvervoltage category
Pollution Degree 2Pollution degree of the in-
tended environment
IP21Degrees of protection
provided by enclosure
Front of instrument (the operation panel side): 70 cmFree space required
around
Amersham Typhoon
scanner
10 cmRight
10 cmLeft
10 cmRear
10 cmTop
A stable laboratory bench with a load capacity of 2940
N (300 kg), or higher.
The bench must be level within two degrees.
Placement
134 Amersham Typhoon Operating Instructions 29193226 AC
9 Reference information
9.1 Specifications