8.1 General errors and warnings
If an error occurs
When an error message appears in the Amersham Typhoon Control Software, follow
the steps below.
ActionStep
Take note of the error code and error message on the monitor.1
Turn off the Amersham Typhoon scanner and the computer, wait 10 seconds
and turn them on again.
2
Try to perform the action again. If the error persists, contact your GE repre-
sentative.
3
General error messages
Corrective actionMeaningError message
The software Help text requires
an installed web browser on the
computer used to control the
scanner.
Failed to open Help in software.
Free up disk space on the
computer.
Store the data on a different
disk.
The available disk space is insuf-
ficient.
The disk capacity is insufficient.
Please free memory on the hard
disk to scan.
Check the computer require-
ments, see Computer require-
ments, on page 84.
This message is displayed when
the Amersham Typhoon Control
Software cannot collect the data
from the memory in the scanner.
Scan stopped. Please check
performance of this PC.
128 Amersham Typhoon Operating Instructions 29193226 AC
8 Troubleshooting
8.1 General errors and warnings