5.1 Site requirements
Precautions
CAUTION
Install the Amersham Typhoon scanner in a location where it will
not come into contact with water or chemicals.
NOTICE
The Amersham Typhoon is intended for indoor use only.
Space requirements
RequirementParameter
110 cm × 90 cm (including required free space at
the left, right and rear)
Required bench space area
70 cm at the front of the instrumentFree space required around
Amersham Typhoon scanner
10 cm at left, right, top and rear of the instrument
Stable, horizontal surface.Placement
Do not place in direct sunlight or in brightly lit
places.
82 Amersham Typhoon Operating Instructions 29193226 AC
5 Installation
5.1 Site requirements