Dimensions
The following illustration shows the dimensions of Amersham Typhoon scanner.
90 cm
80 cm
40 cm
Amersham Typhoon scanner weighs 90 kg.
Environmental requirements
RequirementParameter
18°C to 28°CAmbient temperature, operation
20% to 70%, non-condensingMax. relative humidity, operation
2000 m above sea levelMax. altitude, operation
Pollution Degree 2Pollution degree of the intended
environment
Electrical requirements
RequirementParameter
100 to 240 V AC ±10%Supply voltage
50 to 60 HzFrequency
2.5 to 1.0 AMax current
Overvoltage category IITransient overvoltages
Amersham Typhoon Operating Instructions 29193226 AC 83
5 Installation
5.1 Site requirements