Revision D 170 Series Monitor 8-3
2003023-001
Cleaning: Tocotransducer, Ultrasound Transducer, and Legplate
Tocotransducer, Ultrasound Transducer, and
Legplate
CAUTIONS
ABRASIONāDo not use abrasive cloth, sharp objects, or
abrasive cleaners.
ALCOHOLāDo not use Alcohol in cleaning solutions.
DISCONNECTIONāDetach the transducers/cables/legplate
from the monitor.
IMMERSIONāDo not immerse transducers/cables/legplate or
hold under running water.
1. Dampen a cloth or paper towel with one of the following products; then wring
out until only slightly wet:
ī Sodium Hypochlorite 5.25 % (Bleach) diluted 10:1
ī Cidex
ī Sporicidin
ī Soap and water
2. Rub soiled area until clean, taking care not to excessively wet the
tocotransducer diaphragm seal or the contact surface of the ultrasound
transducer.
3. Use a cloth dampened with water on the contact surface of the ultrasound
transducer and around the seal of the tocotransducer. Do not use a sharp object
which might damage the seal of the tocotransducer.
4. Dry all accessories with a soft, dry cloth.