GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 14 Direction 18000
4-3-59
TABLE 4-11-2
ODC MODULE ERROR CODES
CODE TEST POSSIBLE FAILURE
1 RESERVED
2 DYNAMIC RAM TEST Data read from the dynamic RAM did not compare with data written.
3 BMC DATA BUFFER Data read from the BMC data buffer did not compare with data
written.
4 DUAL PORT RAM Data read from dual port ram did not compare with data written.
5 5380 SCSI IC Data read from 5380 register did not compare with data written, or a
forced 5380 interrupt was not detected.
6 DYNAMIC RAM
PARITY
DRAM parity error interrupt not successfully forced or cleared.
7 NOT USED
8 NOT USED
Operational Errors:
Some error conditions may occur while the board is executing M level programs. These errors will by
reported on the Left RED LED. The error codes are listed in Table 4-11-3. The error can be cleared by
Rebooting the Computer.
TABLE 4-11-2
ODC MODULE ERROR CODES
CODE ERROR
9 BMC Data Buffer Parity Error
10 Dynamic RAM Parity Error
11 Illegal Interrupt From ARZ Side
12 Unable to issue Mode Select to ODD (Optical Drive)
Note that Codes 9, 10, & 11 indicate possible hardware problems on the controller board. Code 12 can be
caused by simply having the ODD turned “OFF”.