GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 1 Direction 18000
6-10-1
SECTION 10 ALIGNMENT LIGHT ADJUSTMENT
10-1 Introduction
DANGER: LASER RADIATION WHEN OPEN AVOID DIRECT EXPOSURE TO BEAM TO PREVENT
EYE INJURY.
CAUTION: Laser radiation do not stare into beam.
WARNING: LASER APERTURE DO NOT STARE INTO BEAM.
The only adjustment to be made by servicemen is the alignment adjustment. Any malfunction of the laser lights
requires replacement only.
NOTE: Prior to performing the Alignment light adjustment, the X-Ray plane of rotation adjustment must be
correct.
LASER WARNINGS AND PRECAUTIONS
ILLUSTRATION 6-10-1