GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-10-17
7. CTDS Diagnostics Loading
To load the diagnostic programs onto the system operating Disk pack, mount and load Diagnostics.
Load-from-Warm Proprietary diagnostic tape onto the tape drive. Tape Drive must be in HIGH Density.
R
DIR DA0 ‘CR’ (Argus/Ace) DIR DZ0 ‘CR’ (Zebra)
R
UPSYS ‘CR
R
INIT MT0 ‘CR
R
MT0:0 ‘CR’
(type <ctnrl>shift><p> simultaneously when instructed)
Copyright GE 1990, all rights reserved
Licensed Material. Property of GE MSBG
Please wait, Initialization in progress...
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
! 9800 CT/T Proprietary Diagnostics !
! LOAD PROCEDURE !
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
...
...
...
...
...
...
DAS SELF TEST SUCCESSFUL
DAS Type / Configuration identified as : (das type) / (detector type)
Done
[ MT0 RELEASED ]
[ LOAD COMPLETED]
[ FOR RELEASE INFORMATION TYPE CTDS.RN ]
STRIKE ANY KEY TO CONTINUE <-----Wait for this message to appear before pressing a key.
NOTE: To view the Release Notes for the CTDS software, proceed as follows:
TYPE CTDS.RN (In CTDS Directory).
Use Control S to stop scrolling and Control Q to continue.