GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-10-19
M -------------- Is the actual CT 9800 Scan-Display-Reconstruction program
TAPES ---------- 8800-to-9800 archive-tape conversion program
EMIARCHIVE -8800 program. Use 8800 ARCHIVE and REMOVE also
DOWN ----------- Allows orderly shutdown of the operating environment
CLEANUP -------- Allows cleanup of DATABASE if problems. -use sparingly-
............. Also special programs for Calibration
QCAL ----------- This is a special calibration required for new tubes
REGEN ---------- This program establishes the CT values of phantoms
and allows modification of selected INSTALL parameters
............. Also special programs for Characterization and Status
MG (CTDS) ----- This test program ‘Characterizes’ the XRAY Generator
CH (CTDS) ----- This program ‘Characterizes’ the Table mechanics
ZR (CTDS) ----- PPI status display program
.....Pick your option terminated with ‘RETURN’ or ‘ENTER’
R
9. Restore All Customer data Files:
a. Load the Scan Rx patient default files, air and phantom calibration files, table/gantry characterization, x-
ray calibration and image tape record from tape to disk. This procedure will be dependent upon which
method was used in paragraph 1 to save the customer data files.
If the SAVEALL program was available for use in paragraph 1, the RESTORALL program may be used
now. In this case proceed to subparagraph b below; otherwise proceed with subparagraph c below.
b. Use the RESTORALL program to restore the Rx patient defaults files, air and phantom calibration files,
table/gantry characterization files, and the x-ray calibration and image tape record from the SAVEALL
tape produced under paragraph 1.
Mount the SAVEALL tape on the tape drive prior to proceeding:
R
DIR DA0 ‘CR’ (Argus) DIR DZ0 ‘CR’ (Zebra)
R
RESTORALL ‘CR’
RESTORING 9800 CUSTOMER FILES
PLEASE MOUNT A ‘SAVEALL’ TAPE
This procedure will complete in about 2 minutes
****** PRESS CONTROL-A TO CANCEL OR: ******
STRIKE ANY KEY TO CONTINUE
...
... (The program will display on the screen those files being restored from tape to disk.)
...
R
NOTE: If there is a problem in loading the files after entering RESTORALL, load the data from
the secondary backup file as follows:
R
DIR DA0 ‘CR’ (Argus) DIR DZ0 ‘CR’ (Zebra)
R
RESTORALL2 ‘CR’