GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 1 Direction 18000
6-3-21
3-4 Disk System Adjustments (Continued)
3. At the DTOS prompt (==>), observe the following, and type the underlined responses:
==> LOAD ZDKP RELI ‘CR’
LOAD:
ZDKP RELI REV. nn
STARTING ADDRESS = 505
SET SWPAK AS PER 8.0, HIT ’CR’ TO CONT.
4. At the S/140 front console, hold the UNLOCK switch, and press RESET.
5. At the Virtual Console prompt (!), observe the following and type the underlined responses:
!11A nnnnnn 040000 ’LF’ (Set Switch
Register Switch 1)
!503R (Restart at 503)
DCH BLK SIZE -10 ’CR’
UNIT: 0 ’CR’ (Dual Drive Systems- 0 Primary Drive)
- 1 Secondary Drive
DATA: ’CR’
COMMAND STRING: SEEK 0 SEEK 1376 LOOP ’CR’
6. To terminate the continuous seeks, hold the UNLOCK switch and press RESET at the S/140
front console.