EasyManua.ls Logo

GE CT-9800

GE CT-9800
510 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 19 Direction 18000
6-9A-14
15) Turn the mounting studs the specified amount to bring the detector into alignment. One turn of the
mounting adjustment moves the detector 0.071” (1.81mm)
Note: The detector adjustment hardware works such that CW adjustment moves the detector
toward the gantry. This is the opposite of Xenon.
16) While holding the adjustment stud with a 7/16” wrench, re-tighten the adjustment locking nut.
17) Torque the inside detector mounting nuts to 25 ft-lbs. When the inside nuts have been torqued, torque
the outside nuts to 25 ft-lbs while holding the inside nut from turning. If a torque wrench is not available
the nuts should be tightened finger tight and then tightened 1/2 flat with a wrench. Tighten the large
detector mounting nuts to 25 ft-lbs.
18) Repeat steps 4) thru 11) until the detector is within tolerance.
9A-7 REFERENCE & DATA CHANNEL OUTPUT check
and Tube X-RAY OUTPUT (DOSE MEASUREMENT) check
Perform a reference and Data channel output check at the following times:
a) New installation: Perform anytime after KV/mA calibration, but before Q-Cal intensity check.
b) Tube replacement: Perform anytime after KV/mA calibration, but before Q-Cal intensity check.
c) Preliminary check when System Dose measurements are questioned.
d) Check for bad channels
In addition: Perform a Tube X-Ray output (Dose Measurement) check at the following times:
e) When the System Dose measurements are questioned.
f) To check X-ray Intensity when the High Voltage subsystem output is suspect.
These procedures assume that the following items have been verified:
a) There must be nothing obstructing the X-ray beam
b) There is nothing blocking any detector channels (except for the Q-CAL channel).
c) Check the tube housing for the correct filter:
1 dot = titanium - the correct filter for XENON detector
2 dots = molybdenum - the correct filter for HILIGHT detector
d) Check the tube
Generator calibration at 80 KVP/40 ma is accurate to within 1KV and 1ma
Generator 40ma station is within spec.
Bleeder output is 80KV
e) Check alignments:
POR - Plane of Rotation alignment
BOW - Beam on Window alignment
Aperture (collimator) spacing is exact ( or on the wide side of the specification)
X-ray calibration
f) Check that the detector has no electrical shorts or opens.
and for XENON:
- The bias supply is 350VDC loaded with less than 10 mV P=P ripple
- The pressure is within specification

Table of Contents

Related product manuals