GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-3-13
To set a breakpoint, the following procedure is used:
From DIR DA0 (or DZ0) go to “M” level and hit the “MANAGE” key to display this menu:
___________________________________________________
MANAGE MENU
SPECIFY SELECTION => _6_ Enter option 6
1 = REMOVE
2 = TUBE WARM-UP
3 = CALIBRATION
4 = SETUP FILES
5 = SYSTEM STATUS
6 = SERVICE SUPPORT
___________________________________________________
MANAGE MENU
SERVICE SUPPORT
SPECIFY SELECTION => _2_ Enter option 2
1 = EXAMINE ERROR LOG
2 = TERMINATE/START ERROR LOG
3 = EXAMINE SCAN LOG
4 = N/A
5 = N/A
6 = EXAMINE PROCESS STATUS