GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-4-7
4. Then we will Read Switch Register and
force a ’PANIC’ by typing ------------------------------------
! !
11/ ! (PANIC HANDLER ADDR) ! <LF>
------------------------------------
I Type a capital I to reset the S140.
(PANIC HANDLER ADDR.)R Type the panic handler address followed by a capital R.
The System should respond with a message indicating a ’CPU malfunction...’
P Type a capital P to proceed
UNIT 3 READY? CPU message tells you to mount a tape on the drive and set the tape drive
select switch to unit #3 (thumbwheel switch).
P Type a capital P to proceed. The CPU then writes an image of the core
memory to tape.
DONE When CPU is done, remove the tape and set the tape drive select switch to
unit # 0.
5. Set the tape drive thumbswitch back to 0.
6. Reset the CPU, Bring the disk drive to ”READY. Bring the system to the ’R’ level.
7. Enter INIT MT0
8. Enter DIR EDD;DUMP MT0:1 CTEX.SW CTSTAT
9. Enter RELEASE MT0
10. Dismount the tape from the tape drive.
11. Please send the tape, a copy of the recorded information and any additional comments to :
GE Medical Systems
CT Service Engineering (NB918)
16800 W. Ryerson Rd.
New Berlin, WI 53151