GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-6-16
*** Deadstart Program (DS) *** Revision 1.05 *** NO DASIOP BOARD ***
Configuration Information
Processor Type = ST-X
01 Compute Head (s)
02 Back Projector board(s)
ACU Rounding Mode = ST-100
SMP MUX NOT PRESENT
Main Memory 1.0 Mega words
*** RESET Complete ***
PASSED Master ROM test
PASSED Master private RAM test
PASSED Master public memory test
PASSED Master TIMER test
PASSED Slave CP test
PASSED Slave ROM test
PASSED Slave private RAM test
PASSED Slave public memory test
*** RESET Complete ***
Deadstart Part 2 Download in Progress
........
PASSED IOP0 test
PASSED PLS Head 0 test
PASSED Backprojector Controller test
Main memory = 1.0 Mega words found
IOP’s are = --- CHI ---
*** Deadstarts Complete ***
*** Deadstart Program (DS) *** Revision 1.05 *** NO SMP BOARD ***
Configuration Information
Processor Type = ST-X
01 Compute Head (s)
02 Back Projector board(s)
ACU Rounding Mode = ST-100
SMP MUX NOT PRESENT
Main Memory 1.0 Mega words
*** RESET Complete ***
PASSED Master ROM test
PASSED Master private RAM test
PASSED Master public memory test
PASSED Master TIMER test
*** FAILED Slave CP test
*** FAILED Slave ROM test
*E* *** FAILED Slave private RAM test, failing
address = 0000
*E* *** FAILED Slave public memory test, failing
address = 0000
PASSED IOP0 test
*** FAILED PLS Head 0 test
*** FAILED Backprojector Controller test
failed Back Projectors
00 01 02 03 04 05
Strings which failed the Master Clear test
SMP0 ACP0
Main memory = 1.0 Mega words found
IOP’s are = DAS CHI ---
Deadstarts Complete ***

Table of Contents