GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-6-30
503 01F7 MPIOUTL CONFIG. FILE ERROR, MASTER, PM
Insufficient memory for short IORP’s (master).
504 01F8 MPIOUTL CONFIG. FILE ERROR, MASTER, PM
Insufficient memory for long IORP’s (master).
505 01F9 MPIOUTL CONFIG. FILE ERROR, MASTER, PM
Insufficient memory for buffers (master).
506 01FA MPIOUTL CONFIG. FILE ERROR, SLAVE, PM, MASTER
Insufficient memory for short IORP’s (slave).
507 01FB MPIOUTL CONFIG. FILE ERROR, SLAVE, PM, MASTER
Insufficient memory for long IORP’s (slave).
508 01FC MPIOUTL CONFIG. FILE ERROR, SLAVE, PM, MASTER
Insufficient memory for buffers (slave).
600 0258 MPVACM BAD APM LOAD, MASTER
Allocate IORP error - insufficient memory available system will have to be
reconfigured with less IOPs.
601 0259 MPVACM MASTER, PM
Allocate device control block extension error - same cause and corrective
action as above (600).
602 025A MPVACM WRONG IOP DEVICE, BAD IOP(DAS,HOST)
Device type error - device doesn’t agree with MG$IOPAT edit SYSAT.SA to show
correct AP configuration or repair device.
700 02BC MPIBMCM BAD APM LOAD, MASTER
Insufficient memory available - you have configured too many IOPs.
701 02BD MPIBMCM - NOT USED -
IOP-1 is not an IBM IOP.
702 02BE MPIBMCM - NOT USED -
IOP-2 is not an IBM IOP.
703 02BF MPIBMCM - NOT USED -
IOP-3 is not an IBM IOP.
704 02C0 MPIBMCM - NOT USED -
IOP-4 is not an IBM IOP.
705 02C1 MPIBMCM - NOT USED -
IOP-5 is not an IBM IOP.
706 02C2 MPIBMCM - NOT USED -
IOP-6 is not an IBM IOP.
707 02C3 MPIBMCM - NOT USED -
IOP-7 is not an IBM IOP.
710 02C6 MPIBMCM - NOT USED -
NO memory for DCBX.
711 02C7 MPIBMCM - NOT USED -
No memory for IORP.
712 02C8 MPIBMCM - NOT USED -
No memory for CCB.
713 02C9 MPIBMCM IOP, IOC, PM, MMC
IOP-1 not responding.
714 02CA MPIBMCM IOP, IOC, PM, MMC
IOP-2 not responding.
715 02CB MPIBMCM IOP, IOC, PM, MMC
IOP-3 not responding.
716 02CC MPIBMCM IOP, IOC, PM, MMC
IOP-4 not responding.