GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-3-49
THE FOLLOWING IS THE CONFIGURATION INFORMATION FOR AN HTD SYSTEM WITH DUAL ACE
DRIVES:
Enter command (? to see choices): A<CR>
You must have a printer at device code 17 to log your dialogue to a printer.
Do you have such a printer (Y or [N]) ? N<CR>
?
Disk Selection
CT - Department CT C1 - Department CT using 1 ST34520N
HTD - High Throughput Dynamic UD - User-defined
ND - No Disk
Select Disk (or EN or END disk selection): HTD<CR>
Enter BMC priority (0=lowest, 7=highest: S280 and MV4000 0-3 only) [5 ]: <CR>
...Field accepted.
What is the throttle burst rate [32 ] ? <CR>
What is the throttle break count [4 ] ? <CR>
Current Configuration Facts:
Logical Logical Logical Logical
Unit 0 Unit 1 Unit 2 Unit 3
Heads 1 1 1 1
Sectors 128 128 128 128
Cylinders 8538 611 0 20
Formt’d megs 559 39 0 1
Logical blks 1092614 78001 0 2337
To see the partitioning of a logical unit, enter the unit number: <CR>
______________________________________________________________________
Microcode on logical unit 0
BMC priority ............ 5
Throttle burst rate ..... 32
Break count ............. 4
Enter command (? to see choices): ?U<CR>
Updating...Please wait.
...EPROM updated successfully.
If you have changed something and do not wish to have the changes saved use the “[Q]uit ” command.
Enter command (? to see choices): Q<CR>
EEPROM not updated! Do you really want to abandon these edits (Y/N) ? Y<CR>
After the program has terminated, you may need to do some of the
following:
Format or reformat the disk.
Initialize or re-initialize the disk.
Do you want to configure another GESCSI controller (Y/[N]) ?N<CR>
!
MR - Department MR
M1 - Department MR using 1 ST34520N