Chapter 3. Logging Data
58 DigitalFlow™ GF868 Programming Manual (1-Channel)
3.5.3 START DATE Prompt
11. Press [F1] to accept the displayed Start Date or press [F2] to enter a different start date. To start logging today, press
[F3].
• If OK or TODAY was selected, you have finished setting up the error log. Proceed to Procedure Options below.
• Otherwise, complete the following steps to Edit the Start Date.
a. Enter the desired Year and press
[ENT]. The acceptable range is 0 to 99.
b. Use the [, [ and [F1]-[F4] keys to select the desired Month.
c. Enter the desired Day and press [ENT]. The acceptable range is 1 to the number of days in the selected month
(28, 29, 30 or 31).
3.5.3a Procedure Options
You have completed setting up the error log. Do one of the following:
• Use [ and [ to review the previous menu prompts.
• Press [F1]-[F4] to select one of the LOG submenus.
• To return to the data screen and begin logging data, press the [ENT] key. The * on the locator bar indicates that the
Model GF868 is now compiling the specified error log.
The error log will continue to run until it is manually stopped, the meter runs out of memory (for a non-circular log), or
the entire 120 records (2 pages x 60 records/page) have been logged.