Chapter 1. Calibration
14 DigitalFlow™ GS868 Service Manual
1.8.3 Entering Pulse Number
5. Enter the number of pulses desired (between 1 and 10,000) and press the [ENT] key. That number of pulses will then
be output at the specified frequency.
6. Repeat Steps 3, 4 and 5 to test all four of the frequency/totalizer outputs. If any of the outputs fails to pass the test,
contact the factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of the following:
To calibrate/test additional slots, press [F4] (EXIT) and proceed to the appropriate section.
To leave the Calibration Menu, press [F4] (EXIT) twice.