1.12 Leak Test
To carry out leak tests, proceed as follows:
Press the key once and scroll to the 'Leak'
function using the key. Press key to select.
Presented with the leak test sub menu, the options 'Start' and
'Set Time' are given as shown below.
To set the duration of the leak test.
Scroll using the key to select 'Set Time' and press the
key.
With the first digit now flashing, press and then select
the number required using the keys and press
key. ( e.g. time set to 999 min.)
With the second digit flashing, press and again select
the number required and press key.
Repeat procedure for third number and press key.
Now select the units of time measurement ( e.g. seconds,
minutes, hours, days, etc. ) Scroll through the units with the
keys and press
Once the leak test time has been set press the key to
return to the leak test sub-menu.
Select 'Start' using the keys and press
to start the leak test.
The top line shows 0.0000
pressure and the bottom line
shows the current pressure.
Start the leak test by pressing the key.
The top line displays the difference in pressure while the
bottom line shows the start pressure and the time counting
down.
When the leak test has finished, to start the leak test again
press the key once to reset the time and again to
start the test.
If no other leak test is to be carried out, press the key
to return to the leak test sub-menu.
Leak Sub-menu
1.12.1 Set Time
MENU
MENU
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Page 17
K0354, Iss. 1