QUICK REFERENCE
Keys
Press to turn PC6-IDOS on and off
Press to access the menus
Press to accept functions or settings
Press to move cursor up/left or increase values
Press to move cursor down/right or decrease values
Press to print/log/start event depending on previous
setting
Select menu position cursor press the Enter key
under function
To exit out of any menu select 'Exit' option,
or press key.
Activating Functions
Exit
Menu Selection
Pressure
Display
1st Press for Menu 1
2nd Press for Menu 2
3rd Press back to Pressure Display /
Calibration Menu (see page 33)
Press or to position cursor and press
to select required alphanumeric. Press
key to release cursor.
For next data entry, repeat the above procedures then
press once all data is entered.
Note: To quickly scroll through alphanumeric lists hold
down the or key.
Data Entry
Tare
This is a
temporary function and will be cancelled when the unit
is switched off.
Hold down the key for 4 seconds while in
the pressure display mode. The unit will bleep after two
seconds when the Max. and Min. values reset. After
another two seconds a second bleep will sound as the
displayed pressure now resets to zero.