For absolute transducers it is possible to adjust the pressure
reading at barometric pressure. This function should only be
performed when there is access to a high accuracy barometer,
and both the barometer’s and ’s pressure sensing
To adjust the barometric reading of an absolute transducer,
proceed as follows:
elect the zero function as described on the previous
page.
You will now be presented with the current barometric
pressure reading. When the pressure has stabilised,
press the key.
• Press the key twice.
• Scroll to 'Power_Off' function using the key and press
the key.
• Select 'Key' to turn the auto-power function off.
• Select 'Timed' to set the auto-power off time (in minutes)
and press the key. This automatically turns the auto-
power function on.
• Set to desired time using the keys and press the
key.
PC6-IDOS
ports must be at the same height. The pressure reading can
only be altered by up to 5% of Full Scale pressure.
• S
•
• The upper line display is the current pressure reading;
the lower line display is the desired pressure reading.
• Adjust the lower line pressure value using the
keys and the key. Once you are satisfied with the
pressure value, press the key.
• Provided that the desired change is acceptable, the
adjustment will be performed, and you will see the
message:
This function allows the user to select the current (displayed)
pressure module.
• Press the key twice to select the Set Menu.
• Scroll to 'Port' using the key and press the key
• Select a module using the keys and press the
key to activate. When selecting a different
pressure module, the serial number and range of the
module will be displayed briefly before the unit returns to
pressure display.
2.11 Port
2.12 Power_Off
The PC6-IDOS has an auto-power off function. The time after
which the unit automatically switches off is adjustable from 1 to
999 minutes. However, the auto-power off will not operate under
the following conditions:
If a key has been pressed during the auto-power off
time period.
If a leak test is running.
If the RS232 Status is On.
If the Log Status is On.
If the unit is online with SiCal PRO.
±
K0354, Iss. 1