EasyManua.ls Logo

GE HumiLab - Chamber response time

GE HumiLab
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 — Operation Page 17
Front panel displays and switches
Refer to Figure 7 as you review the following descriptions of the
front panel displays and selector switches.
The Humilab, as shipped, is set up by the factory for normal opera-
tion. The two upper digital displays read Percent RH and Tempera-
ture in °C. The lower digital display shows the desired chamber set
point.With the exception of the setpoint selector switch, the other
front panel switches are seldom used in day-to-day operation.
However, they are included in case you wish to make changes to
normal default operation and for system testing.
Appendix B contains information on changing factory-set defaults.
A. HUMIDITY Display. This display is shipped from the factory set
for percent relative humidity. Please refer to Appendix B for more
information.
Figure 7 — Front Panel Controls

Table of Contents

Related product manuals