Chapter 2 — Operation Page 21
Chamber response time
Due to the large cavity size, typical response time will be approxi-
mately 30–40 minutes. At the end of that period, the RH in the test
chamber will be within 1% RH of the set point. After that time, the
two will track each other within the accuracy specification of the
instrument.