Page 42 Chapter 5 — Maintenance and Troubleshooting
If that is possible, run it again with a stable water supply.
Relative humidity is a function of temperature. Are you cali-
brating devices in the chamber that generate heat?
If you are, try to locate the heat producer adjacent to the temper-
ature sensor in the center of the chamber. This way, the system
will correct for it automatically.
Is the green Control Status indicator off?
If so, the chilled mirror is not tracking the chamber. See the sec-
tion on page 59 in Appendix B on chilled mirror hygrometer
system troubleshooting.