4 COMPLIANCE
The product is designed to meet the following type tests as listed in Table 4-1.
Table 4-1: Type Tests
CATEGORY
STANDARD
Performance
Criteria
TEST
EMC Emissions
EN 61326-
1:2006
CISPR 11
A
Radiated & Conducted Emissions
FCC Part 15
Meets the
requirements of A
Radiated & Conducted Emissions
EN 61000-3-2
A
Harmonic Current Emissions
Limits
EMC Immunity
EN 61326-
1:2006
IEC 61000-6-5:
2015
EN 61000-4-2
B
Electrostatic Discharge
EN 61000-4-3
A
Electromagnetic Field Immunity
EN 61000-4-4
B
Electrical Fast Transients
EN 61000-4-5
B
Surge Immunity
EN 61000-4-6
A
Conducted RF Immunity
EN 61000-4-8
A
Magnetic Field Immunity
EN 61000-4-11
B, C
Voltage Dips & Interruptions
IEC 61000-4-16
A
Mains frequency voltage
IEC61000-4-17
A
Ripple on DC power supply
IEC61000-4-18
B
Damped oscillatory wave
EN 60255-5
5 kV, 2 kV &
500 V DC
Impulse, Dielectric & Insulation
resistance testing
Environmental
Tests
IEC 60068-2-1
−40 °C
Cold
IEC 60068-2-2
55 °C
Dry Heat
IEC 60068-2-6
10 500 Hz,
0.5 g operation
10 500 Hz,
1 g endurance
Vibration
IEC 6IEC0068-2-
30
55 °C, 95% RH
Damp Heat
IEC60068-2-6
Vibration (sinusoidal)
IEC60068-2-27
Vibration (bump, shock)
EN 60529
IP56
Degree of Protection
Safety
IEC 61010-1
2010
EN 61010-1
2010
UL61010-1
*5
CSA C22.2 No.
61010-1
*5
*5
Note: Bushing monitoring falls outside the scope of UL61010-1 and CSA C22.2 No.
61010-1 and cannot be included in any third-party certification for North
America.