Loading...
GE
Medical
Systems
MAC
1200
V
1.2
Page
48
Information
Technologies
227
492
21
Rev
F
No
display
on
the
screen
Does
the
yellow
Stop
LED
on
the
keypad
glow
after the
device
is
switched
on?
>
if
not:
-
refer
to
"Device
cannot
be
switched
on
....”
>
if
yes:
-
continue
-
Does
a
beep
sound
approx.
10s
after
switching
on
the
MAC1200
occur
(indicates
the
successful
power
up
self-test)?
>
if
not:
-
Self
test
error,
PCB
Control
defective
>
ifyes:
-
continue
-
Contrast
badly
adjusted?
-
Can
ithe
background
illumination
be
activated
by
pressing
a
key?
-
Can
the
display
test
in
Section
3.3
be
applied
successfully?
>
ifnot:
-
graphics
display
module
defective
-
or
PCB
Control
defective
Error
in
self-test
identified
When
an
error
is
detected
during
the
self-test,
in
addition
to
the
message
“Self-test
failed”,
the
error
code
number
and
a
short
description
also
appear
on
the
display.
The
meaning
of
the
error
codes
is
described
in
Section
3.7.
The
error
codes
refer
to
the
PCB
Control,
and
should
be
noted
as
information
for
the
service
center.
MAC
1200
faiis
to
give
printout,
no
paper
transport
Perform
the
following
test
in
the
6-lead
operating
mode:
-
Paper
available?
Paper
correctly
inserted?
Paper
transport
problems
(jam)?
-
Paper
feed
flap
correctly
engaged
on
both
sides?
-.
After
pressing
the
start
key,
the
green
start
Led
must
glow
>
if
not:
-
keypad
defective?
Apply
keypad
test
in
Section
3.4
>
if
yes:
-
continue
-
Paper
feed
flap
correctly
engaged
on
both
sides?
-
Mark
reader
defective
or
not
plugged
in?
-
All
connections
for
printhead
and
Motor
are
plugged
in?
-
Motor
blocked
?
(check
roller,
transmission)
-
Operating
only
on
battery
which
is
strongly
depleted?
>
if
yes:
-
plug
in
power
plug
>
if
not:
-
PCB
Control
defective