GE
Medical
Systems
MAC
1200
V
1.2
Page
49
Information
Technologies
227 492
21
Rev
F
Paper
transport
functions,
no
printout
-
Paper
feed
flap
correctly
engaged
on
both
sides?
-
All
connectors
for
printhead
plugged
in?
-
Operation
with
strongly
depleted
battery?
>
if
yes:
-
plug
in
power
plug
>
ifno:
-
PCB
Control
defective
-
or
printhead
defective
MAC
1200
only
prints
on
the
upper
or
lower
section
of
the
printout
-
Paper
feed
flap
is
only
engaged
on
one
side.
MAC
1200
prints,
but
only
baselines
are
printed
out
-
Electrodes
applied
correctly?
-
Electrode
cables
plugged
into
the
patient
trunk
cable
terminal-box
correctly?
-
Patient
trunk
cable
defective
(
e.g.,
RL
defective)?
-
Contact
problems
at
the
patient
input
connector
of
the
MAC1200?
>
if
not:
PCB
Control
defective.