5-20 MAC 5000 resting ECG analysis system Revision B
2024917-010
Troubleshooting: Frequently Asked Questions
Frequently Asked Questions
Maintenance
NOTE
See Operator’s Manual for complete System Setup information.
Save Setups
Q: How do I save changes I have made to the System Setups?
A: Check the following:
Return to Menu by pressing the esc key or selecting More from
the menu until you see System Setup.
Select System Setup.
Select Save Setup.
Select To System.
You can select Main Menu to exit System Setup.
Storing ECGs
Q: Why won't any of the ECGs I perform save to the floppy diskette?
A: Check the following:
Check that the diskette is fully inserted into the drive.
Make sure you are using high-density (HD) 3.5" diskettes.
Verify that the diskette is not write-protected.
Try a new diskette.
If your system is not set up to automatically save records, you
must manually save by pressing Store.
Format a Disk
Q: How do I format a diskette in the MAC 5000?
A: When you insert a new diskette into the MAC 5000 the system will
automatically format the diskette. If you are using a previously-used
diskette, and want to reformat it, you can format it at a PC.
Cleaning
Q: Should I clean the MAC 5000?
A: Clean the exterior surfaces of all the equipment and peripheral
devices monthly, or more frequently if needed.
Use a clean, soft cloth and a mild dishwashing detergent diluted
in water.
Wring the excess water from the cloth. Do NOT drip water or any
liquid on the writer assembly, and avoid contact with open vents,
plugs, and connectors.
Dry the surfaces with a clean cloth or paper towel.