EasyManua.ls Logo

GE MAC 5000

GE MAC 5000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-22 MAC 5000 resting ECG analysis system Revision B
2024917-010
Maintenance: Disassembly Guidelines
Trolley Casters
Removal
1. Remove the MAC 5000 and all loose items from the trolley and then
place the trolley on its side.
2. Locate the slot under the arrow on the bearing dust cap and using a
small blade screwdriver, pry the cap from the caster to be removed.
3. Using an Allen wrench, remove the wheel shaft and wheel from the
caster.
4. Using an Allen wrench, remove the bolt holding the caster to the
trolley.
54A, 55A
56
A
57A

Table of Contents

Other manuals for GE MAC 5000

Related product manuals