Revision B MAC 5000 resting ECG analysis system 4-11
2024917-010
Maintenance: Disassembly Guidelines
7. Gently lift the rear of the top cover free from the unit.
NOTE
The top cover holds the bezel that surrounds the rear panel
connectors, so the bezel may fall free at this time.
8. At the front of the top cover, gently pull the thin strip of plastic free
from under the keyboard. The entire top assembly is now loose.
NOTE
It may be helpful to rotate the top cover 45
° to provide a larger
opening to clear the display.
9. Carefully lift the top assembly up and clear of the raised display.
Reassembly
1. Raise the display to the vertical position.
2. Make sure the bezel surrounding the rear panel connectors is in
place.
3. Lower the top cover down around the display and set in position.
4. Snap the rear of the top cover in place and then, gently pulling on the
thin plastic strip at the front of the top cover, position it in place
under the keyboard assembly.
5. Replace the screws removed in disassembly.