Revision B MACā„¢ 5000 resting ECG analysis system C-5
2024917-010
Appendix C – Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
The MAC 5000 is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure
that the MAC 5000 is used in such an environment.
Immunity Test EN 60601 Test Level Compliance Level Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should not
be used closer to any part of the equipment, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
EN 61000-4-6
3 Vrms
150 KHz to 80 MHz
3 V rms
d = 1.2
Radiated RF
EN 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer, and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with
the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects,
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in
which the equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
equipment.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P

Table of Contents

Other manuals for GE MAC 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5000 and is the answer not in the manual?

GE MAC 5000 Specifications

General IconGeneral
BrandGE
ModelMAC 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals