EasyManua.ls Logo

GE Moisture Image 1 Series - Testing Recorder Outputs

GE Moisture Image 1 Series
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
April 2004
4-10 Programming with PanaView
Testing Recorder
Outputs
1. To test the recorder outputs, click on the [Test] button in the
Recorder Properties” window (see Figure 4-5 on page 4-9). A
message appears: “This will suspend Recorder #[X] on Channel
[X]. Are you sure you want to test this recorder?” Click on [
OK] to
continue.
2. The “Test Recorder” window appears, as shown in Figure 4-6
below. At the top, slide the pointer to the desired percentage of the
full scale recorder range, and click on [
OK].
3. To trim the recorder outputs, scroll the arrows to the right of the
Zero” and “Span” windows up or down until you reach the
desired values, and click on [OK].
Note: For a complete discussion of testing and trimming recorder
outputs, refer to the sections Testing the Recorder Outputs
and Trimming Recorder Outputs in Chapter 2 of the Service
Manual.
The screen returns to the Recorder Properties” window.